Specifications

Operating Specifications*
Range0.03 ppb to 50 ppm
Precision< 1% RSD  
Accuracy± 2% or ± 0.5 ppb, whichever is greater
Display Readout3 significant digits
Sample TypeAutosampler or discrete grab sample
CalibrationTypically stable for 12 months
Analysis Time2 minutes (4 seconds with the optional Turbo mode)
Sample Temperature5–95 °C (41–203 °F)
Ambient Temperature5–40 °C (41–104 °F)
Instrument Sample 
Flow Rate
0.5 mL/min
Analyzer Specifications
OutputsUSB device port (1); USB host ports (3); Modbus TCP/IP
Display7" WVGA 800X480 pixel, Color LCD with touch-screen
Power100 - 240 V~, 50 -60 Hz, 100 VA
FusesReplace with same type and size fuse: T 1.6 A 250 VAC Fuse (Slow Blow), size 5 X 20 mm appliance inlet
DimensionsH: 42.2 cm (16.6 in); W: 24.6 cm (9.7 in); D: 40.0 cm (15.8 in)
Weight9.4 kg (20.6 lb)
Safety CertificationsETL, CE (see Declaration of Conformity
Sample Conductivity (Optional)
Raw Conductivity Range0.01 to 2,000 μS/cm
Conductivity Accuracy±0.005 μS/cm or ±1%.whichever is greater
Precision<0.25% RSD
Environment
Maximum Relative Humidity0 - 95%, non-condensing
Maximum Altitude3,000 m (9.800 ft)
Pollution Degree2
The UV lamp inside this product contains mercury and must be recycled or disposed of in accordance with local, state, and federal laws.
* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.