InnovOx On-Line Specifications

Operating Specifications*
Analysis Modes TC, TIC, NPOC, TOC
Dynamic TOC Range 0.5 to 50,000 ppm
TOC LOD 0.05 ppm
TOC Accuracy ± ≤3% of reading at 50 ppm NPOC
TOC Precision RSD < 5% of reading at 50 ppm NPOC
TOC Linearity R2 > 0.995, measured as NPOC (see Operation & Maintenance Manual for calibration recommendation)
Analysis Time 2.6 to 8.3 min in online mode
TOC Calibration Stability Up to 6 months
Particle Diameters in Sample ≤ 200 µm diameter without filtration, > 200 µm with optional filter panel
Ambient Operating Temperature Range 10-40 °C (50–104°F)
Maximum Relative Humidity Up to 95%, non-condensing
Sample Temperature Range 10-60 °C (50–140 °F)
Sample Inlet Pressure
Up to 125 psig
Drain Gravity drain
Analyzer Specifications
Outputs Ethernet (Modbus and TCP/IP), Binary, 4-20 mA, and USB
Display Color, touch-sensitive LCD
Power 100–240 ± 10% VAC, 280 W, 50/60 Hz
Dimensions H: 92.7 cm (36.5 in); W: 64.6 cm (25.4 in); D: 38.7 cm (15.3 in)
Weight 36.3 kg (80 lb)
Safety Certifications CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2 No. 61010-1. see Declaration of Conformity
IP Rating IP56 (optional)
Consumables***
Acid Reagent As needed
Oxidizer Reagent As needed
Reagent Grade Water As needed
Pumps 24 months
Sample Flow Path Tubing 12 months
Valve Seals (2) Up to 3 months, depending on sample matrix
Sample Pump Tubing (optional) 3 months

* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.