InnovOx On-Line Specifications
| Operating Specifications* | |
|---|---|
| Analysis Modes | TC, TIC, NPOC, TOC |
| Dynamic TOC Range | 0.5 to 50,000 ppm |
| TOC LOD | 0.05 ppm |
| TOC Accuracy | ± ≤3% of reading at 50 ppm NPOC |
| TOC Precision | RSD < 5% of reading at 50 ppm NPOC |
| TOC Linearity | R2 > 0.995, measured as NPOC (see Operation & Maintenance Manual for calibration recommendation) |
| Analysis Time | 2.6 to 8.3 min in online mode |
| TOC Calibration Stability | Up to 6 months |
| Particle Diameters in Sample | ≤ 200 µm diameter without filtration, > 200 µm with optional filter panel |
| Ambient Operating Temperature Range | 10-40 °C (50–104°F) |
| Maximum Relative Humidity | Up to 95%, non-condensing |
| Sample Temperature Range | 10-60 °C (50–140 °F) |
| Sample Inlet Pressure |
Up to 125 psig |
| Drain | Gravity drain |
| Analyzer Specifications | |
| Outputs | Ethernet (Modbus and TCP/IP), Binary, 4-20 mA, and USB |
| Display | Color, touch-sensitive LCD |
| Power | 100–240 ± 10% VAC, 280 W, 50/60 Hz |
| Dimensions | H: 92.7 cm (36.5 in); W: 64.6 cm (25.4 in); D: 38.7 cm (15.3 in) |
| Weight | 36.3 kg (80 lb) |
| Safety Certifications | CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2 No. 61010-1. see Declaration of Conformity |
| IP Rating | IP56 (optional) |
| Consumables*** | |
| Acid Reagent | As needed |
| Oxidizer Reagent | As needed |
| Reagent Grade Water | As needed |
| Pumps | 24 months |
| Sample Flow Path Tubing | 12 months |
| Valve Seals (2) | Up to 3 months, depending on sample matrix |
| Sample Pump Tubing (optional) | 3 months |
* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.