The 900 Portable is available in Turbo mode for fast response applications.
900 Portable Turbo Specifications
The Sievers 900 Portable Turbo provides the fastest, most accurate total organic carbon (TOC) data across the widest range of any analyzer. Specifications include an analysis time of 4 seconds, a response time of 2.5 minutes, and a wide analytical range of 0.20 parts per billion (ppb) to 10 parts per million (ppm) of TOC. The Sievers 900 On-Line Turbo is:
- Sensitive enough to detect the smallest changes in reclaim TOC
- Robust enought to maintain accuracy even at 10 ppm
- Fast enough to outperform all other reclaim TOC analyzers and provide up-to-the-second results
- Immune to false positives and false negatives that trouble non-membrane-based TOC analyzers
| Operating Specifications1 | |
|---|---|
| Range | 0.20 ppb to 10 ppm |
| Precision | 2% RSD |
| Accuracy | ± 3% |
| Display Readout | 3 significant digits |
| Sample Type | On-line continuous |
| Calibration | Typically stable for 12 months |
| Analysis Time | 4 seconds |
| t52 | Time to detect 5% of TOC step change |
| t902 | Time to detect 90% of TOC step change |
| Sample Temperature2 | 1–95 °C (34–203 °F) |
| Ambient Temperature | 10–40 °C (50–104 °F) |
| Sample Pressure3 | Up to 250 psi |
| Sample Flow Rate | 30–300 mL/min (for on-line mode) |
| Instrument Sample Flow Rate |
1.1. ml/min; 0.5 mL/min when TOC >5 ppm |
| Analyzer Specifications | |
| Outputs | 4-20mA output (1); Serial RS-232 output (2); USB port (1); parallel printer port (1) |
| Display | Quarter-VGA, color touch-screen display |
| Power | Universal Power Supply: 100–240 ± 10% VAC, 50 W, 50/60 Hz |
| Dimensions | H: 35.6 cm (14.0 in); W: 22.3 cm (8.8 in); D: 46.5 cm (18.3 in) |
| Weight | 12.5 kg (27.5 lb) |
| Safety Certifications | ETL, CE |
| Consumables | |
| UV Lamp | 6 months |
| Acid Reagent | As needed, typically 6 months (285 mL) |
| Oxidizer Reagent | As needed, typically 3-month stability; available in 150- or 300-mL cartridge |
1 Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
2 When TOC is less than 5 ppm.
3 If the sample temperature and pressure are above 60 °C and 100 psi, the 900 Analyzer model with the stainless steel iOS is required.