900 Portable Specifications
| Operating Specifications1 | |
|---|---|
| Range | 0.03 ppb to 50 ppm |
| Precision | < 1% RSD |
| Accuracy | ± 2% or ± 0.5 ppb, whichever is greater |
| Display Readout | 3 significant digits |
| Sample Type | On-line continuous, Autosampler, or discrete grab sample |
| Calibration | Typically stable for 12 months |
| Analysis Time | 4 minutes (4 seconds for the optional Turbo mode) |
| Sample Temperature2 | 1–95 °C (34–203 °F) |
| Ambient Temperature | 10–40 °C (50–104 °F) |
| Sample Pressure2 | Up to 250 psi |
| Sample Flow Rate | 50–300 mL/min (for on-line mode) |
| Instrument Sample Flow Rate |
0.5 mL/min |
| Analyzer Specifications | |
| Outputs | 4-20mA output (1); Serial RS-232 output (2); USB port (1); parallel printer port (1) |
| Display | Quarter-VGA, color touch-screen display |
| Power | Universal Power Supply: 100–240 ± 10% VAC, 50 W, 50/60 Hz |
| Dimensions | H: 35.6 cm (14.0 in); W: 22.3 cm (8.8 in); D: 46.5 cm (18.3 in) |
| Weight | 12.5 kg (27.5 lb) |
| Safety Certifications | ETL, CE see Declaration of Conformity |
| Consumables | |
| UV Lamp | 6 months |
| Acid Reagent | As needed, typically 6 months (285 mL) |
| Oxidizer Reagent | As needed, typically 3-month stability; available in 150- or 300-mL cartridge |
1 Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
2 If the sample temperature and pressure are above 60 °C and 100 psi, the 900 Analyzer model with the stainless steel iOS is required.