System suitability, calibration, verification and validation standards for the Sievers 500 RL.

500 RL Standards

•  Annual Kits
•  System Suitability
•  Calibration
•  Verification
•  Validation

 Part Number
 Description

Annual Kits

500 RL with Super iOS

STD KIT5MS Annual Pharmaceutical Standards Kit for Monthly Testing
 

To order this kit, use the following part numbers and quantities:

12 x CSTD 74503 (System Suitability Set) - Shipped one per month.  System Suitability Set includes one Reagent Water Control (Rw), one Standard Solution (Rs), and one System Suitability Solution (Rss).  These standards are packaged in a vial cartridge for use with the Sievers Super iOS System.

STD KIT5QS Annual Pharmaceutical Standards Kit for Quarterly Testing
 

To order this kit, use the following part numbers and quantities:

4 x CSTD 74503 (System Suitability Set) - Shipped one per quarter.  System Suitability Set includes one Reagent Water Control (Rw), one Standard Solution (Rs), and one System Suitability Solution (Rss).  These standards are packaged in a vial cartridge for use with the Sievers Super iOS System.

STD KIT5WS Annual Pharmaceutical Standards Kit for Weeky Testing
 

To order this kit, use the following part numbers and quantities:

52 x CSTD 74503 (System Suitability Set) - Two sets shipped every two weeks.  System Suitability Set includes one Reagent Water Control (Rw), one Standard Solution (Rs), and one System Suitability Solution (Rss).  These standards are packaged in a vial cartridge for use with the Sievers Super iOS System.

500 RL with Standard iOS

STD KIT5M Annual Pharmaceutical Standards Kit for Monthly Testing
 

To order this kit, use the following part numbers and quantities:

12 x CSTD 74403 (System Suitability Set) - Shipped one per month.  System Suitability Set includes one Reagent Water Control (Rw), one Standard Solution (Rs), and one System Suitability Solution (Rss).

STD KIT5Q Annual Pharmaceutical Standards Kit for Quarterly Testing
 

To order this kit, use the following part numbers and quantities:

4 x CSTD 74403 (System Suitability Set) - Shipped one per quarter.  System Suitability Set includes one Reagent Water Control (Rw), one Standard Solution (Rs), and one System Suitability Solution (Rss).

STD KIT5W Annual Pharmaceutical Standards Kit for Weekly Testing
 

To order this kit, use the following part numbers and quantities:

52 x CSTD 74403 (System Suitability Set) - Two sets shipped every two weeks.  System Suitability Set includes one Reagent Water Control (Rw), one Standard Solution (Rs), and one System Suitability Solution (Rss).

System Suitability and Compendial Standards

500 RL with Super iOS

CSTD 74503 500 Super iOS System Suitability
 

Includes one Reagent Water Blank - Rw, one Standard Solution - Rs (500 µg C/L as USP Sucrose), one Standard Solution - Rss (500 µg C/L as USP 1.4-Benzoquinone), and one Rinse Water Standard. These standards are package in a vial set cartridge for use with the Sievers Super iOS System.

500 RL with Standard iOS

CSTD 74403 500 System Suitability Set
 

Includes one Reagent Water Blank - Rw, one Standard Solution - Rs (500 µg C/L as USP Sucrose), one Standard Solution - Rss (500 µg C/L as USP 1,4-Benzoquinone), and one Rinse Water Standard.

Japanese Pharmacopoeia Testing

CSTD 74529-01 500 RL with Super iOS Japanese (JP) Protocol Set (TOC)
 

This standard set meets the new Japanese Pharmacopoeia TOC requirements and includes: (2) 30-mL plastic vial standards, (1) reagent water blank, (1) 500-µg C/L as SDBS standard, and (1) empty glass vial for alignment.

Calibration Standards

500 RL with Super iOS

CSTD 74501 500 Super iOS Single Point Cal ibration Set
 

Includes two Calibration Blanks, one 1.5 mg C/L TOC Calibration Standard (as KHP) and one 25 µS/cm Conductivity Standard (as HCI). These standards are packaged in a vial set cartridge for use with the Sievers Super iOS System.

CSTD 74505 500 Super iOS Single Point Calibration - TOC Only
 

Includes two Calibration Blanks and one 1.5 mg C/L TOC Calibration Standard (as KHP) packaged in a vial set cartridge for use with the Sievers Super iOS System. Not compatible with 500 RL instruments equipped with the conductivity option.

CSTD 74512 500 Super iOS Single Point Calibration/Verification
 

Includes: one Single Point Calibration Set and one Accuracy/Precision & Verification Set. These standards are packaged in a vial set cartridge for use with the Sievers Super iOS System.

CSTD 74514 500 Super iOS Multi-Point Calibration Set
 

Includes three Calibration Blanks, three TOC Calibration Standards (500 µg C/L, 1.0 mg C/L, 1.5 mg C/L as KHP) and one 25 µS/cm Conductivity Standard (as HCI). These standards are package in a vial set cartridge for use with the Sievers Super iOS System.

CSTD 74516 500 Super iOS Multi-Point Calibration/Verification
 

Includes: one Multi-Point Calibration Set and one Accuracy/Precision/Verification Set. These standards are packaged in a vial set cartridge for use with the Sievers Super iOS System.

CSTD 74522 500 Super iOS Single-Point Calibration/Verification - TOC Only
 

Includes one Single Point Calibration Set and one Accuracy/Precision & Verification Set packaged in vial set cartridges for use with the Sievers Super iOS System. Not compatible with 500 RL instruments equipped with the conductivity option.

CSTD 74524 500 Super iOS Multi-Point Calibration Set - TOC Only
 

To order this kit, use the following part numbers and quantities:

52 x CSTD 74503 (System Suitability Set) - Two sets shipped every two weeks.  System Suitability Set includes one Reagent Water Control (Rw), one Standard Solution (Rs), and one System Suitability Solution (Rss).  These standards are packaged in a vial cartridge for use with the Sievers Super iOS System.

CSTD 74526 500 Super iOS Multi-Point Calibration/Verification Set - TOC Only
 

Includes: one Multi-Point TOC Calibration Set and one TOC Accuracy/Precision/Verification Set packaged in vial set cartridges for use with the Sievers Super iOS System. Not compatible with 500 RL instruments equipped with the conductivity option.

500 RL with Standard iOS

CSTD 74401 500 Single-Point Calibration Set
 

Includes two Calibration Blanks, one 1.5 mg C/L TOC Calibration Standard (as KHP) and one 25 µS/cm Conductivity Standard (as HCl).

CSTD 74405 500 Single-Point Calibration Set - TOC Only
 

Includes two Calibration Blanks and one 1.5 mg C/L TOC Calibration Standard (as KHP). Conductivity standards are not included in this set.

CSTD 74612 500 Single-Point Calibration/Verification Set
 

Includes: one Single Point Calibration Set and one Accuracy/Precision & Verification Set.

CSTD 74614 500 Multi-Point Calibration Set
 

Includes three Calibration Blanks, three TOC Calibration Standards ( 500 µg C/L, 1.0 mg C/L, 1.5 mg C/L as KHP) and one 25 µS/cm Conductivity Standard (as HCl).

CSTD 74616 500 Multi-Point Calibration/Verification Set
 

Includes: one Multi-Point Calibration Set and one Accuracy/Precision & Verification Set.

CSTD 74622 500 Single-Point Calibration/Verification Set - TOC Only
 

Includes: one Single Point Calibration Set and one Accuracy/Precision & Verification Set. Conductivity standards are not included in this set.

CSTD 74624 500 Multi-Point Calibration Set - TOC Only
 

Includes three Calibration Blanks, and three TOC Calibration Standards (500 µg C/L, 1.0 mg C/L, 1.5 mg C/L as KHP). Conductivity standards are not included in this set.

CSTD 74626 500 Multi-Point Calibration/Verification Set - TOC Only
 

Includes: one Multi-Point Calibration Set and one Accuracy/Precision & Verification Set. Conductivity standards are not included in this set.

Verification Standards

500 RL with Super iOS

CSTD 74502 500 Super iOS Accuracy/Precision/Verification Set
 

Includes one Verifcation Blank, one 500 µg C/L TOC Verification Standard (as NIST Sucrose) and one 25µS/cm Conductivity Standard (as HCI). These standards are package in a vial set cartridge for use with the Sievers Super iOS System.

CSTD 74507 500 Super iOS Accuracy/Precision/Verification - TOC Only
 

Includes one Verification Blank and one 500 µg C/L TOC Verification Standard (as NIST Sucrose packaged in a vial set cartridge for use with the Sievers Super iOS System. Not compatible with 500 RL instruments equipped with the conductivity option.

500 RL with StandardiOS

CSTD 74402 500 Accuracy/Precision/Verification Set
 

Includes one Verification Blank, one 500 µg C/L TOC Verification Standard (as NIST Sucrose) and one 25 µS/cm Conductivity Standard (as HCl).

CSTD 74407 500 Accuracy/Precision/Verification - TOC Only
 

Includes one Verification Blank and one 500 µg C/L TOC Verification Standard (as NIST Sucrose). Conductivity standards are not included in this set.

Validation Standards

500 RL with Super iOS

CSTD 74506 500 Super iOS Linearity Set
 

Includes, one each 250 µg C/L, 500 µg C/L, 750 µg C/ as KHP and one Linearity Blank. These standards are packaged in a vial set cartridge for use with the Sievers Super iOS System.

CSTD 74513 500 Super iOS Single-Point Validation
 

Includes: one Single Point Calibration Set, two Accuracy/Precision & Verification Sets, one System Suitability Set, and one Linearity Set. These standards are packaged in a vial set cartridge for use with the Sievers Super iOS System.

CSTD 74515 500 Super iOS Multi-Point Validation
 

Includes: one Multi-Point Calibration Set, two Accuracy/Precision & Verification Sets, one System Suitability Set, and one Linearity Set. These standards are packaged in a vial set cartridge for use with the Sievers Super iOS System.

500 RL with Standard iOS

CSTD 74406 500 Linearity Set
 

Includes, one each 250 ug C/L, 500 ug C/L, 750 ug C/L as KHP and one Linearity Blank.

CSTD 74613 500 Single-Point Validation
 

Includes: one Single Point Calibration Set, two Accuracy/Precision & Verification Sets, one System Suitability Set, and one Linearity Set.

CSTD 74615 500 Multi-Point Validation
 

Includes: one Multi-Point Calibration Set, two Accuracy/Precision & Verification Sets, one System Suitability Set, and one Linearity Set.

52 x CSTD 74403 (System Suitability Set) - Two sets shipped every two weeks.  System Suitability Set includes one Reagent Water Control (Rw), one Standard Solution (Rs), and one System Suitability Solution (Rss).