Exhaled Breath Accessories

On-Line NO Breath Kit — Part #CRBK 01400

For measurement of on-line exhaled NO. The kit includes restrictors calibrated to give flows of 30, 50, 100, 150 200 and 250 mL/s BTPS at ~ 10-15 cm H2O mouth pressures.

On-Line NO Breath Kit

ATS Compliant Bag Collection and Sampling Kit — Part #CBSK 01410

This kit includes an inspiratory gas filter to reduce the concentration of NO in inspired air to < 5 ppb and 12 L Mylar® bags to collect the entire vital capacity at a flow rate of 350 mL/s. The bags are equipped with Luer caps to seal the bag after sample collection. The kit includes adapters and three-way valves for analyzing the samples using the NOA.

ATS bag

Dead-space Discard Bag Collection and Sampling Kit — Part #CBSK 01400

The kit includes the inspiratory gas filter and a valve in the body of the filler to permit discard of the first few hundred milliliters of exhaled breath, then collect the remainder of the breath.

Dead space bag

ENO Calibration Gas — Part #CNDK 01400

Calibration of the NOA for gas-phase measurements is easily performed using this lightweight, disposable cylinder of 45 ppm NO in N2.

ENO calibration gas

Zero Air Filter — Part #CACT 01400

Calibration of the NOA for gas measurement also requires a gas contains low levels (< 1 ppb) of NO. This is easily produced from room air using a cartridge containing KMnO4 and activated carbon. The trap has sufficient capacity for several years of use.
 Zero air filter

Calibration Kit — Part #CNCK 01400

Includes the Zero Air Filter, ENO Calibration gas cylinder, a low flow regulator preset to 300 mL/min gas and a calibration tee.
Calibration kit

Bacterial/Viral Filters — Part #CAFL 01420

These single-subject disposable bacterial and viral filters provide > 99.9% filtration efficiency for use with the On-Line NO breath kit and the Bag Collection and Sampling Kits. Package of 50.

 Bacterial/viral filter

Thermal Mass Flowmeter — Part #CACC 01420

On-line measurement of exhaled NO requires exhalation at a constant flow rate. The TSI thermal mass flowmeter permits direct measurement of the exhalation flowrate using the NOA 280i or the NOA 280 equipped with the flowmeter upgrade kit (Part # ACC 01487-01).

The flowmeter is installed on the On-Line NO Breath kit and the signal processed by the NOA. Calibration is performed using standard calibration syringes (1-3 L) and the NOAnalysis 3.21 REB program.

The flowmeter has a fast response time (4 msec) and low pressure drop. The 25.4 mm OD ISO-taper outlet fits over the Accurate NO breath kit mouthpiece and a 22 mm OD ISO-tapered inlet fits the NOA bacterial and viral filters. The polycarbonate body is compatible with standard cold sterilization agents.

The flowmeter employs two sensors; a platinum film sensor is heated and maintained at a constant temperature to measure flow and a temperature sensor is used to compensate for changes in gas temperature.

Specifications
Flow Measurement: Range — 0 - 1.5 L/s BTPS
Accuracy — 3.5% of reading at BTPS
Weight: 180 grams
Dimensions: 182 mm x 64 mm x 53 mm
Operating Temp: 0 to 50 °C

 Thermal mass flowmeter